Specify the Model Number, add the Materials of Construction Code selection to the Model Number and total the prices to arrive at list price.
Product Notes
Material Certificate Not included with the unit. If a Material Certificate is required for SS version, the Batch Number of the material is provided on the quality cert that comes in the box with the sensor.
They can take this batch number and go to the page below and look up the material certificate. http://www.yokogawa.com/eu/an/analytical/eu-material-cert.htm
4-Electrode sensors do not operate with the DC402.
Cell Constants
Sensor Cell Constants MUST match the desired measurement range.
For plug-in type (-*P), use the WU40 cable. For Variopin type (-*V), use the WU10 cable.
Mounting
A FF40 or FS40 is required to properly mount the sensor into the process.
To mount in Immersion, use the FS40 holder and flip it upside down to yield a 1 1/4" NPT thread, where a pipe can be attached.
Conductivity Cells
The Conductivity Cells on this page are the large bore cells 30mm in diameter. See other pages for the 12mm(small bore) series and the 19mm series (SC4A).
Additional Ordering Instructions
Order Construction & Cell Constant in the following combinations: -SP34 2-Electrode Flow Sensor, 316 Stainless Steel Constant = 0.01/cm -SV34 2-Electrode Flow Sensor, 316 Stainless Steel Constant = 0.01/cm -SP24 2-Electrode Flow Sensor, 316 Stainless Steel Constant = 0.1/cm -SV24 2-Electrode Flow Sensor, 316 Stainless Steel Constant = 0.1/cm -EP04 2-Electrode Flow Sensor, Epoxy Constant = 10/cm -EP15 2-Electrode Insertion Sensor, Epoxy Constant = 1.0/cm -EP08 4-Electrode Flow Sensor Epoxy Constant = 10.0/cm (Not for use with DC402) -EP18 4-Electrode Flow Sensor Epoxy Constant = 1.0/cm (Not for use with DC402) -FP08 4-Electrode Flow Sensor PVDF (Kynar) Constant = 10.0/cm (Not for use with DC402) -FV08 4-Electrode Flow Sensor PVDF (Kynar) Constant = 10.0/cm (Not for use with DC402) -TP08 4-Electrode Flow Sensor PTFE (Teflon) Constant = 10.0/cm (Not for with DC402) -TV08 4-Electrode Flow Sensor PTFE (Teflon) Constant = 10.0/cm (Not for with DC402)
How to Order
See additional Note at the bottom for combination possibilities
* (asterisk) Model codes including all * markers are available for quick delivery. Consult factory for more information.
Product Notes
Note Id
Description
1
See "Additional Ordering Instructions". Please note that the -FP04 and -TP04 version was terminated in 2017 and should be replaced with the 4-electrode -FP08 and -TP08.
2
/M Material certificate is no longer called out. The QIC has the heat trace numbers on the certificates that can be pulled for the material certificates in the customer portal.
Add Item to Quote
Please select atleast one item to proceed.
Configuration Error
Selected configuration is not valid. Please read notes and make corrections.
Email Verification Required
Email Verification link has been sent successfully.
Please follow instruction in email to verify